Inject that 1 GHz CW signal into the Port 2 SA RF Input connector on the instrument as illustrated in
From an instrument preset, tune the instrument to 1 GHz, 0 Span. Verify that the input
attenuator (Atten) is set to 10 dB. On the side of the instrument near the RPG tuning knob, there is a
spring door. Open that spring door and locate the IF Out SMB (m) connector. Connect one end of a test
cable to the IF Out connector on the side of the instrument. Connect the other end of the test cable to the
functioning spectrum analyzer. Tune the functioning spectrum analyzer to 33.75 MHz. The I.F. Out
signal level should measure 33.75 MHz at approximately -16 dBm as illustrated in
Figure 4-4.
33.75 MHz I.F. Output Test Setup
Figure 4-5.
33.75 MHz I.F. Output Signal
Since the input attenuator is set to 10 dB, there should be 6 dB of conversion loss from the RF input
connector to the I.F. Out connector.
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...