7-42
N9927-90003
Repair and Replacement Procedures
FieldFox Analyzer
Removing and Replacing the Measurement Group Assemblies
Reinstalling All Measurement Group Assemblies as One Unit
1. Reinstall the I/O side panel, if needed. Refer to
“Replacing the I/O Side Panel, Doors, and Gasket” on
2. Refer to
. Position the measurement group assemblies to the side of the front case.
3. Reconnect the LCD to System board cable at the System board connector and place a strip of
adhesive tape over the connector to secure it. This strip of adhesive tape is provided in the
. If necessary, refer to
“Replacing the A2 LCD, the LCD Shield, and LCD Cables”
4. Position the measurement group assemblies on top of the front case and align the I/O side panel in the
groove in the front case being careful not to damage the gasket.
5. Position (tuck) the LCD to System board cable in between the LCD shield and the A5 System board
(goes between the speaker side panel) so that it will not interfere with the cases when they are
combined.
6. Reconnect the System board to front panel interface board cable at the front panel interface board
connector as shown.
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...