8-6
N9927-90003
General Purpose Maintenance and Service Procedures
FieldFox Analyzer
Firmware Upgrades
Firmware Upgrades
To Check the Current Firmware Revision
•
Press
System(7)
.
•
Press
Service Diagnostics
.
•
Press
System Information
.
The current firmware version is indicated on the “Firmware Release” line.
For information on upgrading FieldFox firmware, go to
www.agilent.com/find/fieldfox2
.
If the firmware upgrade process is interrupted, memory contents in the instrument may become
corrupted. If a problem is suspected, consider doing a fresh installation of the firmware. Follow the
process under
“Install Firmware from SD Card”
“Firmware Related Information”
section of
for details on what happens during the
firmware update process.
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...