N9927-90003
7-45
FieldFox Analyzer
Repair and Replacement Procedures
Post-Repair Procedures
Manual Checks and Procedures
Manual Check of Keyboard
Check that each key works properly. Check that the keyboard backlight LEDs come on when a key is
pressed.
Manual Check of LCD
Press
Service (7) > Display
, then use
Display Colors
and
Brightness
keys to exercise the display.
Check for even brightness in the LCD backlight.
Manual Check of RPG Operation
Activate a marker and use the RPG to move the marker.
Load Instrument Serial Number into the A5 System Board
Instructions for this process may be accessed via the “Load instrument serial number into System Board”
link at
agilent.com/find/fieldfox2
.
Install Firmware from SD Card
A specially configured SD card must be used when the flash memory on the A6 SOM board is new or
when the A6 SOM board memory has become corrupted.
Instructions for creating and using the SD card are available at the “SD card for installing firmware” link
at
agilent.com/find/fieldfox2
.
Compare Serial Numbers
Compare the serial number stored in the instrument memory with the serial number tag on the back of
the instrument.
To access the stored serial number, press:
System (7) > Service Diagnostics > System Information
Check Sound
“Checking Display Brightness and Sound” on page 4-7
.
Check Time and Date
Check the current time and date. Press
System > System Configuration > More > Date and Time
.
It may be necessary to adjust the time zone as well.
See http://na.tm.agilent.com/fieldfox/service/rtc_info.htm
for restrictions and details for setting time and date.
Re-Install Options
See the
“Service Related User Operations”
section of
for information on re-installing options.
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...