7-20
N9927-90003
Repair and Replacement Procedures
FieldFox Analyzer
Removing and Replacing the Case Group Parts
8. Position the rear case directly over, and aligned with, the front case and the side panels as shown.
9. Make sure the battery cable is tucked in between the measurement group assemblies and the speaker
side panel and that it is not in the path of the screw hole indicated.
10. Firmly press the cases together being careful not to damage the gaskets.
11. Install six new screws (item 7) to secure the rear case to the front case. These new screws are
provided in the
.
12. Tighten the six screws (item 8) that secure the RF top panel.
Figure 7-14
Combining the Front and Rear Cases (3)
13. Reinstall the battery. Refer to
“Removing and Replacing the Main Battery” on page 7-6
.
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...