N9927-90003
8-5
FieldFox Analyzer
General Purpose Maintenance and Service Procedures
Options
Options
Installed options determine the functionality of the analyzer. There is a license associated with each
option.
To View Currently Installed Options
•
Press
System(7)
.
•
Press
System
Configuration
.
•
Press
Options
.
•
Press
Show
Options
.
A list of the currently installed options is displayed.
Re-Installing Options
If an instrument looses all of its options (i.e. it reports or “shows” no options), the only functionality
available will be the base mode for the model.
When the A6 SOM board is replaced, it is necessary to re-load the license files because the license and
option information is stored on the A6 SOM board.
Process for obtaining and re-installing options:
1. Send an email to
with the following information.
•
Indicate that you need previously issued licensed for a FieldFox repair.
•
Model number and serial number of the instrument.
•
Maximum frequency of the instrument from the label above the display.
•
A list of options from the serial number label on the back of the instrument.
•
A brief description of why options need to be re-installed. This information will help Agilent
identify recurring problems that may need special attention.
2. Within two working days, you will receive an email response with a license file attached.
3. Copy the license file to the root of a USB flash drive. Follow the appropriate instructions in the
N9912A Upgrade Installation Instructions (N9912-90005) to install the licenses. These instructions
are available by searching for “N9912-90005” on the Web at
www.agilent.com
.
Additional Information About Options
For information on acquiring additional options, go to
www.agilent.com/find/fieldfox2
and follow the
“
Upgrades available for FieldFox N9912A
” link. There is an installation note associated with each
option upgrade that provides detailed instructions. Some combinations of options are not valid; see the
web information for details.
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...