N9927-90003
5-19
FieldFox Analyzer
Theory of Operation
Grounding and Power
Power Control and Management
Power States
The On/Off Button is used to switch between On, Off, and Standby power states:
Off
Only the System Monitor and battery control circuits are powered up (if power is available from
the battery or external power).
If the battery is charging, the LED below the On/Off Button slowly flashes orange with increasing
intensity, otherwise this LED is off.
Some battery power is required in the “Off” state. A fully charged battery will discharge completely
in four to six weeks when left in the analyzer.
On
This is normal operation. The On/Off Button LED is steadily lit green.
Standby
Only the digital power supplies are on. The CPU is in a low power state, maintaining RAM
content. The On/Off Button LED flashes yellow briefly every two seconds.
During normal operation, unexpectedly disconnecting power to the analyzer (such as removing the main
battery when the analyzer is not connected to external power) should not cause any problems. Simply
re-apply power. No firmware problems are expected. Of course, this does not apply during a firmware
upgrade.
Smart Power Management
The analyzer uses extensive “load switching” to minimize power dissipation. This is termed Smart
Power Management (SPM). There are two motivations for power management:
•
to extend battery run time
•
to prevent overheating.
Sections of the circuitry are selectively powered down via “load switching”, depending on the operating
mode of the analyzer. Five different sections of the circuitry on the A4 RF board may be controlled via
“load switching”. These sections are as follows:
•
Source—switched off during SA mode, switched off between sweeps during CAT or NA modes.
•
LO—switched off when data is not being taken by receivers.
•
SA—switched off when in one port CAT or NA modes, on all the time in two port mode.
•
Reflectometer—switched to low power through state when RF OUT port receivers are inactive.
•
High-Band Pre-amp—switched off when not required.
On the A5 System board, each of the ADC’s may be individually switched to a low power mode when
not active.
Load switching is generally transparent to the user. One obvious exception is the that the source turns off
during “retrace” in the CAT and NA modes.
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...