Chapter 4 Remote Interface Reference
Using the APPLy Command
73
4
Using the APPLy Command
The
APPLy
command provides the most straightforward method to program
the power supply over the remote interface. You can select the specific
output, output voltage, and output current all in one command.
APPLy
{P6V | P25V | N25V}[,{<
voltage
>| DEF | MIN | MAX}[,{<
current
>| DEF | MIN | MAX}]]
This command is combination of
INSTrument:SELect
, [
SOURce:
]
VOLTage
, and
[SOURce:]CURRent
commands. The values of voltage and
the current of the specified output are changed as soon as the command is
executed.
You can identify each output by the output name (P6V, P25V or N25V) as
described in Table 4-1. For the
voltage
and
current
parameters of the
APPLy
command, the ranges depend on the output currently selected. You can
substitute “MINimum”, “MAXimum”, or “DEFault” in place of a specific
value for the
voltage
and
current
parameters. MIN selects the lowest voltage
and current values allowed for the selected output. MAX selects the highest
voltage and current values allowed. The
default voltage
values are 0 volts
for all outputs. The
default current
values are 5 A for +6V output and 1 A for
±25V outputs. The default voltage and current values are exactly the same as
the
*RST
values.
See Table 4-1 for details of parameters.
If you specify only one value for the parameter, the power supply regards it
as voltage setting value. If you do not specify any value for the parameter,
the
APPLy
command only selects the output specified and acts as the
INSTrument
command.
APPLy? [{P6V | P25V | N25V}]
This command queries the power supply's present
voltage
and
current
values for each output and returns a quoted string. The voltage and current
are returned in sequence as shown in the sample string below (the quotation
marks are returned as part of the string). If any output identifier is not
specified, the voltage and the current of the currently selected output are
returned.
"5.000000,1.000000"
In the above string, the first number 5.000000 is the voltage limit value and
the second number 1.000000 is the current limit value for the specified
output.
Summary of Contents for E3631-90002
Page 9: ...8...
Page 14: ...1 1 General Information...
Page 25: ...24...
Page 26: ...2 2 Initial Operation...
Page 34: ...3 3 Front Panel Operation...
Page 64: ...4 4 Remote Interface Reference...
Page 113: ...112...
Page 114: ...5 5 Error Messages...
Page 124: ...6 6 Application Programs...
Page 137: ...136...
Page 138: ...7 7 Tutorial...
Page 153: ...152...
Page 154: ...8 8 Specifications...
Page 161: ...160...