Chapter 3 Front-Panel Operation
Calibration Overview
61
3
To Change the Security Code
To change the security code, you must
first unsecure the power supply, and then enter a new code.
Be sure to read the security code rules on page 58 before attempting to
secure the power supply.
• Front-Panel Operation:
To change the security code, first make sure that the power supply is
unsecured
. Press the
key after the “CAL MODE” message is
displayed in the calibration mode, enter the new security code using the
knob and resolution selection keys, then press the
key.
Changing the code from the front panel also changes the code required
from the remote interface.
• Remote Interface Operation:
CALibrate:SECure:CODE
<new code>
To change the security code, first unsecure the power supply using the old
security code. Then, enter the new code. For example,
"CAL:SEC:STAT OFF, HP003631"
Unsecure with old code
"CAL:SEC:CODE ZZ001443"
Enter new code
"CAL:SEC:STAT ON, ZZ00143"
Secure with new code
Secure
Secure