Chapter 7 Tutorial
Output Characteristics
144
Figure 7-6. Simplified Diagram of Common Mode and Normal Mode
Sources of Noise
When the load changes very rapidly, as when a relay contact is closed, the
inductance in the hook up wire and in the power supply output will cause a
spike to appear at the load. The spike is a function of the rate of change of
the load current. When very rapid changes in load are expected, a capacitor
with a low series resistance, in parallel with the power supply, and close to
the load is the best way to minimize these voltage spikes.
Summary of Contents for E3631-90002
Page 9: ...8...
Page 14: ...1 1 General Information...
Page 25: ...24...
Page 26: ...2 2 Initial Operation...
Page 34: ...3 3 Front Panel Operation...
Page 64: ...4 4 Remote Interface Reference...
Page 113: ...112...
Page 114: ...5 5 Error Messages...
Page 124: ...6 6 Application Programs...
Page 137: ...136...
Page 138: ...7 7 Tutorial...
Page 153: ...152...
Page 154: ...8 8 Specifications...
Page 161: ...160...