
Chapter 2 Initial Operation
Output Checkout
32
6 Ensure that the current can be adjusted from zero to the maximum
rated value.
Adjust the knob until the ammeter indicates 0 amps and then until the ammeter
indicates 5.0 amps.
7 Check the current function for the +25V supply.
Disable the outputs by pressing the
key and connect a
short across (+) and (COM) output terminals of the ±25V supply with an
insulated test lead. Repeat steps (3) through (6) after selecting the meter and
adjust selection key for the +25V supply.
8 Check the current function for the -25V supply.
Disable the outputs by pressing the
key and connect a
short across (-) and (COM) output terminals of +25V supply with an insulated
test lead. Repeat steps (3) through (6) after selecting the meter and adjust
selection key for the -25V supply.
N o t e
If an error has been detected during the output checkout procedures, the
ERROR
annunciator will turn on. See “Error Messages”, for more
information starting on page 113 in chapter 5
1
You can use the resolution selection keys to move the blinking digit to the right or
left when setting the current.
Output On/Off
Output On/Off
1
+25V
-25V
Summary of Contents for E3631-90002
Page 9: ...8...
Page 14: ...1 1 General Information...
Page 25: ...24...
Page 26: ...2 2 Initial Operation...
Page 34: ...3 3 Front Panel Operation...
Page 64: ...4 4 Remote Interface Reference...
Page 113: ...112...
Page 114: ...5 5 Error Messages...
Page 124: ...6 6 Application Programs...
Page 137: ...136...
Page 138: ...7 7 Tutorial...
Page 153: ...152...
Page 154: ...8 8 Specifications...
Page 161: ...160...