Chapter 5 Error Messages
Execution Errors
115
5
Execution Errors
-101
Invalid character
An invalid character was found in the command string. You may have
inserted a character such as #, $, or % in the command keyword or within a
parameter.
Example:
OUTP:TRAC #ON
-102
Syntax error
Invalid syntax was found in the command string. You may have inserted a
blank space before or after a colon in the command header, or before a
comma.
Example:
VOLT:LEV ,1
-103
Invalid separator
An invalid separator was found in the command string. You may have used
a comma instead of a colon, semicolon, or blank space - or you may have
used a blank space instead of a comma.
Example:
TRIG:SOUR,BUS
or
APPL P6V 1.0 1.0
-104
Data type error
The wrong parameter type was found in the command string. You may have
specified a number where a string was expected, or vice versa.
-105
GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
-108
Parameter not allowed
More parameters were received than expected for the command. You may
have entered an extra parameter, or you added a parameter to a command
that does not accept a parameter.
Example:
APPL? 10
-109
Missing parameter
Fewer parameters were received than expected for the command. You
omitted one or more parameters that are required for this command.
Example:
APPL
Summary of Contents for E3631-90002
Page 9: ...8...
Page 14: ...1 1 General Information...
Page 25: ...24...
Page 26: ...2 2 Initial Operation...
Page 34: ...3 3 Front Panel Operation...
Page 64: ...4 4 Remote Interface Reference...
Page 113: ...112...
Page 114: ...5 5 Error Messages...
Page 124: ...6 6 Application Programs...
Page 137: ...136...
Page 138: ...7 7 Tutorial...
Page 153: ...152...
Page 154: ...8 8 Specifications...
Page 161: ...160...