Chapter 7 Tutorial
Remote Programming
149
7
Remote Programming
During remote programming a constant-voltage regulated power supply is
called upon to change its output voltage rapidly. The most important factor
limiting the speed of output voltage change is the output capacitor and load
resistor.
Figure 7-7. Speed of Response - Programming Up (Full Load)
The equivalent circuit and the nature of the output voltage waveform when
the supply is being programmed upward are shown in Figure 7-7. When the
new output is programmed, the power supply regulator circuit senses that
the output is less than desired and turns on the series regulator to its
maximum value I
L
, the current limit or constant current setting.
This constant current I
L
charges the output capacitor C
O
and load resistor
R
L
parallel. The output therefore rises exponentially with a time constant
R
L
C
L
towards voltage level I
L
R
L
, a value higher than the new output voltage
being programmed.
When this exponential rise reaches the newly programmed voltage level, the
constant voltage amplifier resumes its normal regulating action and holds
the output constant. Thus, the rise time can be determined approximately
using the formula shown in Figure 7-7.
Summary of Contents for E3631-90002
Page 9: ...8...
Page 14: ...1 1 General Information...
Page 25: ...24...
Page 26: ...2 2 Initial Operation...
Page 34: ...3 3 Front Panel Operation...
Page 64: ...4 4 Remote Interface Reference...
Page 113: ...112...
Page 114: ...5 5 Error Messages...
Page 124: ...6 6 Application Programs...
Page 137: ...136...
Page 138: ...7 7 Tutorial...
Page 153: ...152...
Page 154: ...8 8 Specifications...
Page 161: ...160...