114
Error Messages
When the front-panel
ERROR
annunciator turns on, one or more command
syntax or hardware errors have been detected. A record of up to 20 errors is
stored in the power supply's error queue. The power supply beeps once each
time an error is generated.
• Errors are retrieved in first-in-first-out (FIFO) order. The first error returned
is the first error that was stored. When you have read all errors from the
queue, the
ERROR
annunciator turns off.
• If more than 20 errors have occurred, the last error stored in the queue (the
most recent error) is replaced with -350,
“Too many errors”
. No additional
errors are stored until you remove errors from the queue. If no errors have
occurred when you read the error queue, the supply responds with +
0
,
“No
error”
over the remote interface or “NO ERRORS” from the front panel.
• The error queue is cleared when power has been off or after a
*CLS
(clear
status) command has been executed. The
*RST
(reset) command
does not
clear the error queue.
•
Front-panel operation
:
If the
ERROR
annunciator is on, press the
key repeatedly to read
the errors stored in the queue. The error queue is cleared when you read
all errors.
•
Remote interface operation:
SYSTem:ERRor?
Reads one error from
the error queue
Errors have the following format (the error string may contain up to 80
characters
).
-113,
"
Undefined header
"
ERROR
-113
Error
Summary of Contents for E3631-90002
Page 9: ...8...
Page 14: ...1 1 General Information...
Page 25: ...24...
Page 26: ...2 2 Initial Operation...
Page 34: ...3 3 Front Panel Operation...
Page 64: ...4 4 Remote Interface Reference...
Page 113: ...112...
Page 114: ...5 5 Error Messages...
Page 124: ...6 6 Application Programs...
Page 137: ...136...
Page 138: ...7 7 Tutorial...
Page 153: ...152...
Page 154: ...8 8 Specifications...
Page 161: ...160...