24
AIC 5000 Installation and Configuration
AIC 5000 Specifications
Dimensions
Height
3.95 inches
10.3 cm
Width
13.3 inches
33.78 cm
Depth
15.1 inches
38.35 cm
Approximate Weight
21 lb
9.53 kg
Temperature Range
Operating
50° to 95° F
10° to 35° C
Nonoperating
-22° to 140° F
-30° to 60° C
Relative Humidity (noncondensing)
Operating
10–90%
10–90%
Nonoperating (38.7° C max wet bulb)
5–95%
5–95%
Maximum Altitude (unpressurized)
Operating
10,000 ft
3048 m
Nonoperating
30,000 ft
9144 m
Note
Operating temperature is derated 1.0° C per 300 m
(1000 ft) to 3000 m (10,000 ft) above sea level, no
direct sustained sunlight. Maximum rate of change
is 10° C/Hr. The upper limit may be limited by the
type and number of options installed.
Power Supply
115 V
230 V
Operating Voltage Range
90–132 VAC
180–264 VAC
Rated Voltage Range
100–127 VAC
200–240 VAC
Rated Line Frequency
50–60 Hz
50–60 Hz
Power Output
185 W
185 W
Rated Input Current (maximum)
5 A @ 100 VAC
2.5 A @ 200 VAC
Heat Dissipation
Maximum
971 BTU/hr
245 kg-cal/hr
Typical (idle)
256 BTU/hr
65 kg-cal/hr