![Agilent Technologies Agilent 33500 Series User Manual Download Page 198](http://html.mh-extra.com/html/agilent-technologies/agilent-33500-series/agilent-33500-series_user-manual_2867489198.webp)
197
Chapter 3 Features and Functions
Embedded Waveform Editor
3
Full Width
(Gaussian only)
The width of the gaussian bell curve between the
points on the curve that are exactly one-half the bell
curve's height, also called full-width half maximum.
Specified as an integer from 1 to the number of data
points in one cycle (default 10).
Symmetry
(Ramp only)
The percentage of time (per cycle period) that the
ramp is rising. Specified as a decimal number from 0
to 100 (default 100).
Zero Crossing
(Sinc only)
The number of times the waveform crosses the
horizontal axis on one side of the waveform. Specified
as an integer from 0 to 100 (default 100).
Duty Cycle
(Square only)
The percentage of the time (per cycle period) that the
waveform voltage is high. Specified as a percentage
from 0 to 100 (default 10).