![Agilent Technologies Agilent 33500 Series User Manual Download Page 107](http://html.mh-extra.com/html/agilent-technologies/agilent-33500-series/agilent-33500-series_user-manual_2867489107.webp)
106
Chapter 3 Features and Functions
Output Configuration
3
•
Remote Interface Operation
:
OUTPut[1|2] {OFF|ON
}
The
APPLy
command overrides the current setting and automatically
enables the channel output connector.
Waveform Polarity
In the
normal
mode (default), the waveform goes positive during the first
part of the cycle. In the
inverted
mode, the waveform goes negative
during the first part of the cycle.
• As shown in the examples below, the waveform is inverted
relative to
the offset voltage. Any offset voltage present will remain unchanged
when the waveform is inverted.
• When a waveform is inverted, the Sync signal associated with the
waveform
is not
inverted.
•
Front-Panel Operation
: Press
or
or
and then press
Polarity
to toggle between the
Normal
and
Inverted
selections.
•
Remote Interface Operation
:
OUTPut[1|2]:POLarity {NORMal|INVerted
}
No
Offset
Voltage
With Offset Voltage
Summary of Contents for Agilent 33500 Series
Page 13: ...12 ...
Page 16: ...1 1 Quick Start ...
Page 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Page 42: ...2 2 Front Panel Menu Operation ...
Page 90: ...3 3 Features and Functions ...
Page 218: ...4 4 Application Programs ...
Page 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Page 244: ...5 5 Tutorial ...
Page 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Page 262: ...6 Specifications ...