113
Chapter 3 Features and Functions
Pulse Waveforms
3
Edge Times
The edge times set the transition times for the leading and trailing edges
of the pulse, either independently or together. For each transition
(leading or trailing edge), the edge time represents the time between the
10% and 90% thresholds.
• Edge time: 8.4 ns to 1 µs (see restrictions below).
The default edge
time is 10 ns
.
• The specified edge time must fit within the specified pulse width as
shown below. The waveform generator will adjust the edge time as
needed to accommodate the specified pulse width.
•
Front-Panel Operation
: After selecting the pulse function, press the
Edge Times
softkey. Then use the
Lead Edge
and
Trail Edge
softkeys, followed by the knob or numeric keypad to enter the desired
edge times.
•
Remote Interface Operation
:
FUNCtion:PULSe:TRANsition:LEADing
{<
seconds
>|MINimum|MAXimum}
FUNCtion:PULSe:TRANsition:TRAiling
{<
seconds
>|MINimum|MAXimum}
FUNCtion:PULSe:TRANsition[:BOTH]
{<
seconds
>|MINimum|MAXimum}
Summary of Contents for Agilent 33500 Series
Page 13: ...12 ...
Page 16: ...1 1 Quick Start ...
Page 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Page 42: ...2 2 Front Panel Menu Operation ...
Page 90: ...3 3 Features and Functions ...
Page 218: ...4 4 Application Programs ...
Page 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Page 244: ...5 5 Tutorial ...
Page 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Page 262: ...6 Specifications ...