![Agilent Technologies Agilent 33500 Series User Manual Download Page 147](http://html.mh-extra.com/html/agilent-technologies/agilent-33500-series/agilent-33500-series_user-manual_2867489147.webp)
146
Chapter 3 Features and Functions
Frequency Sweep
3
Sweep Mode
You can sweep with either linear or logarithmic spacing. For a
linear
sweep, the waveform generator varies the output frequency linearly
during the sweep. For a
logarithmic
sweep, the waveform generator
varies the output frequency logarithmically.
The selected mode does not affect the sweep return (from stop to start, if
one is set). The sweep return is always linear.
• Sweep mode: Linear (default), Logarithmic, or List.
•
Front-Panel Operation
: After enabling sweeps, press the
Linear
softkey again to toggle between the linear or log mode.
•
Remote Interface Operation
:
[SOURce[1|2]:]SWEep:SPACing {LINear|LOGarithmic}
Sweep Time
The
sweep time
specifies the number of seconds required to sweep from
the start frequency to the stop frequency. The number of discrete
frequency points in the sweep is automatically calculated by the
waveform generator based on the sweep time.
• Sweep time: 1 ms to 250,000 seconds.
The default is 1 second
.
•
Front-Panel Operation
: After enabling sweeps, press
Sweep Time
.
Then use the knob or numeric keypad to enter the desired sweep
time.
•
Remote Interface Operation
:
[SOURce[1|2]:]SWEep:TIME {<
seconds
>|MINimum|MAXimum}
Summary of Contents for Agilent 33500 Series
Page 13: ...12 ...
Page 16: ...1 1 Quick Start ...
Page 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Page 42: ...2 2 Front Panel Menu Operation ...
Page 90: ...3 3 Features and Functions ...
Page 218: ...4 4 Application Programs ...
Page 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Page 244: ...5 5 Tutorial ...
Page 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Page 262: ...6 Specifications ...