![Agilent Technologies Agilent 33500 Series User Manual Download Page 132](http://html.mh-extra.com/html/agilent-technologies/agilent-33500-series/agilent-33500-series_user-manual_2867489132.webp)
131
Chapter 3 Features and Functions
Frequency-Shift Keying (FSK) Modulation
3
FSK Carrier Frequency
The maximum carrier frequency depends on the function selected as
shown below.
The default is 1 kHz for all functions
.
When the
External
source is selected, the output frequency is determined
by the signal level on the rear-panel
Ext Trig
connector. When a logic
low
level is present, the
carrier
frequency is output. When a logic
high
level
is present, the
hop
frequency is output.
• Front-Panel Operation
: To set the carrier frequency, press the
Freq
softkey for the selected function. Then use the knob or numeric
keypad to enter the desired frequency.
• Remote Interface Operation
:
[SOURce[1|2]:]FREQuency {<
frequency
>|MINimum|MAXimum}
You can also use the
APPLy
command to select the function, frequency,
amplitude, and offset with a single command.
Function
Minimum Frequency
Maximum Frequency
Sine
Square
Ramp
Pulse
1
Hz
1
Hz
1
Hz
1
Hz
30 MHz
30 MHz
200 kHz
30 MHz
Summary of Contents for Agilent 33500 Series
Page 13: ...12 ...
Page 16: ...1 1 Quick Start ...
Page 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Page 42: ...2 2 Front Panel Menu Operation ...
Page 90: ...3 3 Features and Functions ...
Page 218: ...4 4 Application Programs ...
Page 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Page 244: ...5 5 Tutorial ...
Page 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Page 262: ...6 Specifications ...