Chapter 3 Features and Functions
Output Configuration
3
Duty Cycle (Square Waves)
The duty cycle of a square wave represents the fraction of time per cycle
that the square wave is at a
high level
(assuming that the waveform is
not inverted).
(See Pulse Waveforms, on page 110 for information about duty cycle for
pulse waveforms.)
• Duty Cycle: 0.01% to 99.99% , at low frequencies; range is reduced
at higher output frequency.
• The duty cycle is stored in
volatile
memory; the default value is 50%.
• The duty cycle setting is remembered when you change from square
wave to another function. The duty cycle setting
does not
apply to a
square waveform used as the
modulating
waveform for AM, FM, PM
or PWM. A 50% duty cycle is always used for a modulating square
waveform. The duty cycle setting applies only to a square waveform
carrier
.
•
Front-Panel Operation
: After selecting the square wave function,
press the
Duty Cycle
softkey. Then use the knob or numeric keypad
to enter the desired duty cycle. If you use the numeric keypad, press
Percent
to complete the entry.
•
Remote Interface Operation
:
FUNCtion:SQUare:DCYCle {<
percent
>|MINimum|MAXimum}
The
APPLy
command automatically sets the duty cycle to 50%.
20% Duty Cycle
80% Duty Cycle
Summary of Contents for Agilent 33500 Series
Page 13: ...12 ...
Page 16: ...1 1 Quick Start ...
Page 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Page 42: ...2 2 Front Panel Menu Operation ...
Page 90: ...3 3 Features and Functions ...
Page 218: ...4 4 Application Programs ...
Page 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Page 244: ...5 5 Tutorial ...
Page 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Page 262: ...6 Specifications ...