138
Chapter 3 Features and Functions
Pulse Width Modulation (PWM)
3
Pulse Waveform
• Pulse is the only waveform shape supported for PWM.
•
Front-Panel Operation
: Press
, then
Pulse
.
•
Remote Interface Operation
:
FUNCtion PULSe
You can also use the
APPLy
command to select the function, frequency,
amplitude, and offset with a single command.
Pulse Period
The range for the pulse period is 33 ns to 1,000,000 s. The default is 100
s.
•
Front-Panel Operation
: After selecting the pulse function, press
and the
Frequency/Period
softkey to toggle to the
Period
setting. Then use the knob or numeric keypad to enter the desired
pulse period. If you use the numeric keypad, press a softkey to specify
the units.
•
Remote Interface Operation
:
FUNCtion:PULSe:PERiod {<
seconds
>|MINimum|MAXimum}