Chapter 3 Features and Functions
Burst Mode
3
Burst Period
The
burst period
defines time from the start of one burst to the start of
the next burst.
Used in the internal triggered burst mode only.
Keep in mind that burst period is different than the “waveform frequency”
which specifies the frequency of the bursted signal.
• Burst period: 1
s to 8000 seconds.
The default is 10 ms
.
• The burst period setting is used only when Immediate triggering is
enabled. The burst period is ignored when manual or external
triggering is enabled (or when the
gated
burst mode is selected).
• It
is not
possible to specify a burst period which is too short for the
waveform generator to output with the specified burst count and
frequency (
see below
). If the burst period is too short, the waveform
generator will automatically adjust it as needed to continuously
re-trigger the burst.
•
Front-Panel Operation
: To set the burst period, press
and
then
the
Burst Period
softkey. Then use the knob or numeric keypad to
enter the period.
•
Remote Interface Operation
:
[SOURce[1|2]:]BURSt:INTernal:PERiod
{<
seconds
>|MINimum|MAXimum}
Summary of Contents for Agilent 33500 Series
Page 13: ...12 ...
Page 16: ...1 1 Quick Start ...
Page 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Page 42: ...2 2 Front Panel Menu Operation ...
Page 90: ...3 3 Features and Functions ...
Page 218: ...4 4 Application Programs ...
Page 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Page 244: ...5 5 Tutorial ...
Page 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Page 262: ...6 Specifications ...