Installation
Agilent 5800 and 5900 ICP-OES User's Guide
27
Parameter
Fully Demountable Torches
Standard
Inert
High solids
Semi-volatile organics
Volatile organics
Construction
Fully
demountable
Fully
demountable
Fully
demountable
Fully demountable
Fully demountable
Injector inner
diameter (mm)
1.8
1.8
2.4
1.4
0.8
Injector material
Quartz
Ceramic
Quartz
Quartz
Quartz
Outer tube length
Medium with
slot
Medium with
slot
Medium with
slot
Medium with slot
Short (no slot)
Outer tube
material
Quartz
Quartz
Quartz
High purity quartz (for
organic solvents)
High purity quartz (for
organic solvents)
Intermediate tube
Quartz tulip
Quartz tulip
Quartz tulip
Quartz tulip
Quartz tulip
Recommended
application
Most sample
types
Hydrofluoric
(HF) acid
digests
Samples with
high total
dissolved solids
Semi-volatile organic
solvents e.g. ketones,
kerosene, Jet-A1
Volatile organic
solvents e.g., gasoline
A fully demountable torch can be adapted for another application by replacing the
injector and/or the outer tube with the configuration recommended for that
sample type. Refer to the Agilent Technologies website for ordering information.
Torch Assembly and Disassembly
Both the semi- and fully demountable torch models allow the outer/intermediate
tube-set to be separated from the base/injector for easier cleaning or replacement.
In addition, the fully demountable torch models allow the injector to be removed
from the base for easier cleaning or replacement.
WARNING
Hot Surface
The torch and torch compartment may become extremely hot during
instrument operation and remain hot so for some time after the plasma has
been switched off. Allow the plasma compartment to cool for at least five
minutes before accessing the plasma compartment. Allow the torch to cool
for two minutes before touching the outer tube or attempting to dismantle
the torch.
Summary of Contents for 5800 ICP-OES
Page 1: ...Agilent 5800 and 5900 ICP OES User s Guide ...
Page 6: ...Contents 6 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left blank ...
Page 38: ...Installation 38 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left ...
Page 50: ...Operation 50 Agilent 5800 and 5900 ICP OES User s Guide This page is intenionally blank ...
Page 65: ......