Installation
Agilent 5800 and 5900 ICP-OES User's Guide
21
All connections of services to the ICP-OES instrument are made on the right side
of the instrument, except for the mains power which is on the left side of the
instrument. Remove the cover on the right side by hand to gain access to the
Ethernet, accessory, water and gas connections. The Power on/off button is
located on the front lower left side of the instrument and the mains power switch
is located on the left back side of the instrument.
Figure 2.
Input and output connections on the side of the ICP-OES instrument
Item
Description
1 Accessory and LAN
cable connections
Agilent accessory connections and shielded Ethernet cable
for PC to instrument communication
2 Argon
gas
inlet
Argon
gas inlet, standard
3 Option gas inlet
Option gas (80% Ar/20% O
2
mix) inlet. Optional nitrogen gas
inlet (not shown)
4 Optics purge filter
Argon or nitrogen gas filter
5 Water outlet
Connect to return from the water chiller
6 Water filter
Coarse particulate water filter
7 Water inlet
Connect to supply from the water chiller
3
5
2
7
1
6
4