Maintenance and Troubleshooting
58
Agilent 5800 and 5900 ICP-OES User's Guide
Figure 15B and C.
Assembled torch cleaning stand with swappable torch plates. Injector soaking in
aqua regia.
Figure 15D.
Quartz outer tube-set soaking in aqua regia.
7
Rinse the torch and outer tube-set thoroughly and dry completely prior to
using the torch in the instrument.
8
Follow the steps outlined below for “Rinsing the torch” and “Drying the torch”
listed below.