Operation
40
Agilent 5800 and 5900 ICP-OES User's Guide
Develop a method
Run samples
Print a report
Turning On the Instrument and Software
Before starting the system, carefully read the Safety practices and hazards section
at the front of this manual and ensure that the laboratory is set up according to the
details specified in the Site Preparation Guide.
Turning on the instrument for the first time (or from long term shutdown)
To turn on the instrument for the first time or from shutdown:
1
Check that the exhaust and External Inlet Duct Adapter (if used) lines are
secured to the ICP-OES instrument, and that the air filter is not blocked.
2
Turn on the laboratory exhaust system.
3
Ensure the gas and water lines are connected to the ICP-OES instrument
4
Turn on the gas supply.
5
Switch on the water chiller.
6
Ensure that the gas and water supplies are turned on and set to the correct
pressures, and that the water chilleris set to the correct temperature.
7
Check that the torch is clean and in good condition and installed with the torch
handle fully closed.
8
Check that all tubing on the spray chamber, nebulizer and peristaltic pump is
installed and correctly connected.
9
Check that the plasma compartment door is fully closed.
10
Check that the Ethernet LAN cable is connected to the computer or Local Area
Network.
11
Switch on the computer, monitor and printer.
12
Plug the ICP-OES cable into the wall socket and set the mains power switch on
the left side of the instrument to ‘On’.
Summary of Contents for 5800 ICP-OES
Page 1: ...Agilent 5800 and 5900 ICP OES User s Guide ...
Page 6: ...Contents 6 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left blank ...
Page 38: ...Installation 38 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left ...
Page 50: ...Operation 50 Agilent 5800 and 5900 ICP OES User s Guide This page is intenionally blank ...
Page 65: ......