Installation
34
Agilent 5800 and 5900 ICP-OES User's Guide
3
To remove the top seal off the quartz tube-set, slide the seal towards the top
end of the tube-set. See Figure 12.
Figure 12.
Removing the top seal off the quartz tube-set
To disassemble the fully demountable torch:
1
Unscrew the locking nut on the base of the torch and remove the injector by
gently pulling on the ball joint.
2
Remove the quartz tube-set from the top of the torch base. If the top seal is
removed with the quartz outer tube-set, slide the top seal off the tube-set
towards the top end.