22
Agilent 5800 and 5900 ICP-OES User's Guide
Instrument Status LED Color Coding
The instrument status indicator light on the front top right of the Agilent ICP-OES
displays different colors to represent the status of the instrument:
A green light indicates the following:
The instrument and software are connected and are ready to attempt a
plasma ignition sequence, or
The instrument and software are connected and the plasma ignition
sequence is underway, or
The instrument and software are connected, the plasma is lit and running
and it is possible to run analysis of samples.
A yellow blinking light indicates the instrument and software are not
connected (instrument is off-line).
A yellow light indicates that the instrument and software are connected, but
the instrument is not ready to begin a plasma ignition sequence due to a
tripped door or torch handle interlock, low gas pressure, low water flow or low
cooling air flow. User intervention is required to resolve the problem.
An orange pulsing (intensifying/fading) light indicates the instrument is
booting up or performing a firmware update.
A red light indicates an instrument firmware error causing a halt to analysis or
restricting normal instrument operation. User intervention is required to
resolve the problem.
Front Power On/Off Button
The LED in the front on/off button indicates the basic instrument on/off state.
Off indicates that the instrument is completely unpowered when the mains
power switch on the left side of the instrument is also off.
Green flash once every 10 seconds indicates most of the instrument is
unpowered, with the exception of the front panel Power on/off button switch
sensing circuit inside the mains power module. Press the front panel Power
on/off button at the front to fully power up the instrument.
Green blinking slow (1 Hz). The mains power module is initializing and running
through power up self-check, or is shutting off the instrument.
Summary of Contents for 5800 ICP-OES
Page 1: ...Agilent 5800 and 5900 ICP OES User s Guide ...
Page 6: ...Contents 6 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left blank ...
Page 38: ...Installation 38 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left ...
Page 50: ...Operation 50 Agilent 5800 and 5900 ICP OES User s Guide This page is intenionally blank ...
Page 65: ......