Installation
32
Agilent 5800 and 5900 ICP-OES User's Guide
Torch disassembly
To disassemble a semi-demountable torch:
You may need a flat head screwdriver to loosen the securing ring if you cannot
gently slide the torch out of the torch body by hand.
CAUTION
Only use the screw driver to loosen the torch securing ring from the torch body if
you are unable to do so by hand.
1
Gently loosen the top seal, either by prodding it using thumb nail (see Figure 9)
or use a wide blade flat screw driver to lever it up. When using a screw driver,
insert the screw driver into the slot as shown in Figure 10, maintaining an up-
down action to lever the seal loose. Repeat the action on the opposite slot.
The seal should be dislodged from the torch body.
Figure 9.
Removing the top seal on a semi-demountable torch by prodding it using a thumb nail