Maintenance and Troubleshooting
Agilent 5800 and 5900 ICP-OES User's Guide
59
Rinsing the torch
To rinse the torch (one piece and semi-demountable torches):
1
Hold the torch with the ball joint connector at the top.
2
Thoroughly flush the inside and outside of the torch with deionized water
(18 M
Ω
.cm) using a wash bottle to direct the water stream. See Figure 16A.
3
Invert the torch (see Figure 16B) so that the quartz tube or injector are at the
top and the ball joint connector is at the bottom. Flush rinse water through the
quartz tubes so that the water flows out of the gas entry ports and ball joint
connector for at least 1 minute.
Figure 16
A. Rinse the injector with deionized water through the ball joint connector.
Summary of Contents for 5800 ICP-OES
Page 1: ...Agilent 5800 and 5900 ICP OES User s Guide ...
Page 6: ...Contents 6 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left blank ...
Page 38: ...Installation 38 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left ...
Page 50: ...Operation 50 Agilent 5800 and 5900 ICP OES User s Guide This page is intenionally blank ...
Page 65: ......