Installation
24
Agilent 5800 and 5900 ICP-OES User's Guide
Connecting ICP Expert software to the ICP-OES
To connect the software and instrument:
1
Start the ICP Expert software by double-clicking on the desktop icon.
2
Click Instrument.
3
Click Connect.
4
Enter the IP address of the instrument or select an existing instrument.
5
Click Connect.
6
Click Close.
Performing a Detector and Wavelength Calibration
To perform a detector and wavelength calibration:
1
Turn on the external exhaust.
2
Start the ICP Expert software and open the Instrument Window.
3
Connect the software to the instrument, if needed.
a
Click Instrument in the ICP Expert toolbar.
b
Click Connect.
c
Select the instrument from the list and the click Connect.
4
Click Calibrate in the ‘Detector’ section.
NOTE
Once the detector calibration is completed, the date and time of the last successful
calibration will be displayed, confirming the detector measurement has been
completed.
5
Ensure a standard glass concentric nebulizer, a double pass spray chamber
and the appropriate plasma torch (for radial or dual view) are installed. Use
manual sampling. The recommended peristaltic pump tubing is White/White
for the pump and Blue/Blue for the drain.
6
Ignite the plasma.
Summary of Contents for 5800 ICP-OES
Page 1: ...Agilent 5800 and 5900 ICP OES User s Guide ...
Page 6: ...Contents 6 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left blank ...
Page 38: ...Installation 38 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left ...
Page 50: ...Operation 50 Agilent 5800 and 5900 ICP OES User s Guide This page is intenionally blank ...
Page 65: ......