Chapter 10
155
Command Reference
:CORR:CKIT:STAN3
1
0
. C
o
mm
and
Refe
re
nc
e
:CORR:CKIT:STAN3
Syntax
[:SENSe]:CORRection:CKIT:STANdard3 <numeric 1>,<numeric 2>
[:SENSe]:CORRection:CKIT:STANdard3?
Description
Defines the values of the LOAD compensation standard for the parameters you specify by
using the
command.
These are set as the standard values for the measurement frequency when executing the
command (set with the
command).
When using the multi-compensation function (set to ON with the
command) with the channel-by-channel definition of the standard values enabled (set to
ON with the
command), the standard values for the selected
channel (selected with the
command) are set at execution of the
command.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands
Equivalent key
sequence
[Load]
(
[Blue]
-
[6]
) -
CorVal
-
RefEnt
<numeric 1>
<numeric 2>
Description
Value of the primary
parameter.
Value of the secondary
parameter.
Range
-999.999 to 999.999
-99.9999E9 to 99.9999E9
Initial value
100E-9
0
Unit
F (farad)
Depends on the type of
parameter.
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Page 288: ...286 Index Index ...