Chapter 10
159
Command Reference
:CORR:DATA
1
0
. C
o
mm
and
Refe
re
nc
e
•
When specifying STANdard3 as parameter 1:
In any case, if the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if the upper
limit of the range is exceeded) is set.
MAX or MIN can be used to specify parameter 2 and parameter 3.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands
Equivalent key
sequence
[Open]
(
[Blue]
-
[4]
) -
MeasVal
[Short]
(
[Blue]
-
[5]
) -
MeasVal
[Load]
(
[Blue]
-
[6]
) -
CorVal
-
MeasVal
You can only check the setup values. You cannot set them.
Parameter 2: <numeric 1>
Parameter 3: <numeric 2>
Description
Value of the primary parameter
*1
.
Value of the secondary
parameter
.
Range
-999.999 to 999.999
-99.9999E9 to 99.9999E9
Initial value
100E-9
0
Unit
F (farad)
Depends on the type of parameter.