196
Chapter 10
Command Reference
:TRIG:SOUR
:TRIG:SOUR
Syntax
:TRIGger[:SEQuence1]:SOURce {INTernal|MANual|EXTernal|BUS}
:TRIGger[:SEQuence1]:SOURce?
Description
Selects the trigger mode from the following four types.
Internal
Configures the instrument to use its internal trigger source so that is it
automatically and continuously triggered.
Manual
Configures the instrument to be triggered when you press the
[Trig]
key on the front panel.
External
Configures the instrument to be triggered when a trigger signal is
inputted through the Ext TRIG terminal or handler/scanner interface.
GPIB (Bus)
Configures the instrument to be triggered when the
command is
executed.
Parameters
Query response
{INT|MAN|EXT|BUS}<newline><^END>
Related commands
Equivalent key
sequence
[Trig Mode]
Note that you cannot set the trigger mode to GPIB (Bus) using the front panel keys.
Description
INTernal (initial value)
Specifies Internal.
MANual
Specifies Manual.
EXTernal
Specifies External.
BUS
Specifies GPIB (Bus).