Chapter 3
31
Setting Up Measurement Conditions and Display
Sample Program
3
. S
e
tti
ng u
p
M
eas
ur
em
ent
Con
d
itio
n an
d Di
sp
lay
Sample Program
Example 3-1 shows a sample program to set up the measurement conditions and display.
This program is stored on the sample program disk under the filename “setup.bas.”
This program performs reset, sets up the 4288A as shown in the following table, and saves
the setups into the built-in nonvolatile memory (save location number: 9).
Item
Setup
Measurement
parameter
Primary parameter
Cp
Secondary parameter
D
Measurement signal
Frequency
1 kHz
Level
500 mV
Measurement range
Auto range mode
Measurement time
Long mode
Averaging count
4
Cable length
0 m
Trigger delay time
1 ms
Display ON/OFF
ON
Number of display digits for the measurement
result
5
Fixed point display for the measurement result
OFF
Deviation
measurement mode
(Primary parameter)
Mode
Percentage display
Reference value
10 nF
Deviation
measurement mode
(Secondary
parameter)
Mode
OFF
Reference value
0
Page number of the instrument setup display area
4 (the page that displays the measurement signal level
monitor result)
Measurement signal
level monitor
Current
OFF
Voltage
ON
Beep mode
FAIL
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Page 288: ...286 Index Index ...