Chapter 10
195
Command Reference
:TRIG
1
0
. C
o
mm
and
Refe
re
nc
e
:TRIG
Syntax
:TRIGger[:SEQuence1][:IMMediate]
Description
Immediately generates a trigger and executes a measurement regardless of the setup of the
trigger mode.
If the trigger system is not in the trigger wait state (trigger event detection state), executing
this command causes an error and the command is ignored.
For details on the trigger system, refer to “Trigger system” on page 52. (No query)
Equivalent key
sequence
No equivalent key is available on the front panel.
:TRIG:DEL
Syntax
:TRIGger[:SEQuence1]:DELay <numeric>[MS|M|S]
:TRIGger[:SEQuence1]:DELay?
Description
Specifies the waiting time between when a trigger is inputted and when the measurement
starts (trigger delay time).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set. Note that a fraction below the resolution is rounded off.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Equivalent key
sequence
[Delay]
(
[Blue]
-
[Trig Mode]
)
<numeric>
Description
Trigger delay time
Range
0 to 1
Initial value
0
Unit
s (second)
Resolution
1E-3
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Page 288: ...286 Index Index ...