Displaying measurement result as deviation from reference value
(deviation measurement mode)
You can use the deviation measurement mode to display a relative measurement result as
deviation from the reference value instead of displaying its absolute value. To turn
ON/OFF the deviation measurement mode, use the following commands.
•
•
You can select from two modes in the deviation measurement mode: displaying the
deviation as it is or displaying the deviation as a percentage relative to the reference value.
To set the deviation measurement mode, use the following commands.
•
•
To set the reference value in the deviation measurement mode, use the following
command.
•
The table below shows the relationship between the setups using the above commands and
the values displayed as the measurement result.
Where,
and
are:
: Measured value
: Reference value (set using the
command)
NOTE
The measured value read out by the GPIB command is a calculation result based on the
above setup. For judgment in the comparator function, the measurement result is used as is
regardless of the setup. (Refer to the data processing flow in the
Operation Manual
.)
Setup of
or
Setup of
or
Value displayed as the
measurement result
OFF
———
ON
DEV
PCNT
Meas
Meas
Ref
–
Meas
Ref
–
Ref
----------------------------
100
×
Meas
Ref
Meas
Ref
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Page 288: ...286 Index Index ...