40
Chapter 4
Preparation for Accurate Measurement (Executing Compensation)
Executing OPEN/SHORT/LOAD Compensation
Cable, F_shift, Open, Shor, Load, Load_ref, and Load_para$ to the
file.
Line 1520
Displays the name of the file to which the compensation state is saved.
The Inp_data sub-program of Lines 1540 to 1680 used to input data is described below.
Line 1590
Makes the setup to return to the entry start line when an error (for
example, due to improper input) occurs so that the user can make the
entry again.
Lines 1610 to 1620 Prompts the user to enter the data value specified with Mes$ and waits
for the user to enter the value.
Lines 1630 to 1650 Displays the entered value and waits for the user to enter yes or no
(the y/n key).
Line 1660
If a key other than the y key is pressed in Line 1650, returns to the
entry start line (Line 1610).
Example 4-1
Executing the OPEN/SHORT/LOAD compensation (compen.bas)
10
DIM File_name$[20]
20
INTEGER Result
30
CLEAR SCREEN
40
!
50
ASSIGN @Agt4288a TO 717
60
!
70
File_name$="Cmp_data"
80
!
90
OUTPUT @Agt4288a;":SYST:PRES"
100
!
110
Result=FNCompen(@Agt4288a,"Open",.00002)
120
IF Result<>0 THEN Prog_end
130
Result=FNCompen(@Agt4288a,"Short",20)
140
IF Result<>0 THEN Prog_end
150
Result=FNCompen(@Agt4288a,"Load",.2)
160
IF Result<>0 THEN Prog_end
170
!
180
CALL Save_comp_data(@Agt4288a,File_name$)
190
!
200 Prog_end:
END
210
!=============================================
220
! Compensation Data Measurement Function
230
!=============================================
240
DEF FNCompen(@Agt4288a,Standard$,Limit)
250
DIM Inp_char$[9],Buff$[9],Std$[9],Err$[50]
260
REAL Curr_freq,Freq(1:2),Para1,Para2,Zm,Ym,Gm,Bm
270
REAL Cpref,Dref,Zref,Gref,Bref
280
INTEGER Err_flag
290
OUTPUT @Agt4288a;":SOUR:FREQ?"
300
ENTER @Agt4288a;Curr_freq
310
Freq(1)=1.E+3
320
Freq(2)=1.E+6
330
Load_para$="CPD"
340
OUTPUT @Agt4288a;":STAT:OPER:ENAB 128"
350
OUTPUT @Agt4288a;"*SRE 128"
360
SELECT Standard$
370
CASE "Open"
380
Std$="STAN1"
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...