Chapter 10
147
Command Reference
:CALC1:FORM
1
0
. C
o
mm
and
Refe
re
nc
e
:CALC1:FORM
Syntax
:CALCulate1:FORMat {CP|CS}
:CALCulate1:FORMat?
Description
Specifies the primary parameter to be measured.
When the secondary parameter has been set to G or RP and the primary parameter is set to
CS, the secondary parameter is automatically changed to D. Also, when the secondary
parameter has been set to RS and the primary parameter is set to CP, the secondary
parameter is automatically changed to D.
Parameters
Query response
{CP|CS}<newline><^END>
Related commands
Equivalent key
sequence
[Meas Prmtr]
:CALC1:MATH:EXPR:CAT?
Syntax
:CALCulate1:MATH:EXPRession:CATalog?
Description
Reads out available parameters used when executing the command to specify the
expression of the primary parameter in the deviation measurement mode (the
(
command). The query response is always DEV,PCNT.
(Query only)
Query response
DEV,PCNT<newline><^END> (fixed)
Related commands
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
CP (initial value)
Specifies the capacitance value measured using the parallel
equivalent circuit model for the primary parameter.
CS
Specifies the capacitance value measured using the series
equivalent circuit model for the primary parameter.
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Page 288: ...286 Index Index ...