270
Appendix F
Error Messages
Error number: -220
-220
Parameter error
An error not included in the error numbers between -221 and -229 has occurred during the
analysis of a program data element. This error occurs, for example, when you attempt to
specify invalid values (values not finite when converted to an R-X format impedance
value) as the LOAD correction data or LOAD correction reference data. If this error
occurs, the command is ignored. This error also occurs when you attempt to specify an
invalid LOAD correction reference value by using the front panel keys.
-221
Setting conflict
A program data element complying with the syntax standard has been analyzed, but the
4288A cannot execute it at present.
-222
Data out of range
A data element (that does not violate the standard) has been received out of the range
defined for the 4288A.
-223
Too much data
The received block, equation, or string type program data complies with the standard, but
the amount of data exceeds the limit that the 4288A can handle due to memory or
device-specific conditions related to memory.
-230
Data corrupt or stale
The data is invalid or a newly initiated read operation has not been completed since the
latest access.
-241
Hardware missing
The received command or Query complied with the standard but could not be executed due
to hardware-related reasons (for example, an option was not installed).
Parameter Type Setting
Values not finite when converted to an R-X
format impedance value
Primary
Parameter
Secondary
Parameter
Cp
D
Independent of D value, Cp is 0
Cp
Q
Independent of Q value, Cp is 0, also
Independent of Cp value, Q is 0
Cp
G
Cp and G are both 0
Cp
Rp
Independent of Cp value, Rp is 0
Cs
D
Independent of D value, Cs is 0
Cs
Q
Independent of Q value, Cs is 0, also
Q is 0 independent of Cs value
Cs
Rs
Cs is 0 independent of Rs value
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Page 288: ...286 Index Index ...