160
Chapter 10
Command Reference
:CORR:LOAD
:CORR:LOAD
Syntax
[:SENSe]:CORRection:LOAD[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:LOAD[:STATe]?
Description
Turns ON/OFF the LOAD compensation function.
With the LOAD compensation set to ON, if you change the cable length (set with the
command) or frequency shift (set with the
command), the
LOAD compensation is automatically changed to OFF.
Parameters
Query response
{1|0}<newline><^END>
Related commands
Equivalent key
sequence
[Load]
(
[Blue]
-
[6]
) -
On/Off
Description
ON or 1
Turns ON the LOAD compensation.
OFF or 0 (initial value)
Turns OFF the LOAD compensation.
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Page 288: ...286 Index Index ...