Chapter 1 Quick Start
To Configure a Channel for Scanning
To Configure a Channel for Scanning
Any channel that can be “read” by the instrument can also be included in
a scan. This includes readings on multiplexer channels, a read of a
digital port, or a read of the count on a totalizer channel. Automated
scanning
is not
allowed with the RF multiplexer, matrix, actuator,
digital output, or voltage output (DAC) modules.
1 Select the channel to be added to the scan list.
Turn the knob until the desired channel is shown on the right side of
front-panel display. The channel number is a three-digit number; the
left-most digit represents the slot number (
1
00,
2
00, or
3
00) and the two
digits on the right indicate the channel number (1
02
, 1
10
, etc.).
Note:
You can use
and
to skip to the beginning of the previous or
next slot.
For this example, assume that you have the 34901A multiplexer
installed in slot 100 and select channel 103.
2 Select the measurement parameters for the selected channel.
Use the knob to scroll through the measurement choices on each level of
the menu. When you press
to make your selection, the menu
automatically guides you through all relevant choices to configure a
measurement on the selected function. When you have finished
configuring the parameters, you are automatically exited from the menu.
The current selection (or default) is displayed in full bright for easy
identification. When you make a different selection, the new choice is
shown in full bright and it becomes the default selection. The order of the
choices always remains the same; however, you always enter the menu
at the current (full bright) setting for each parameter.
Note:
The menu will time-out after about 20 seconds of inactivity and any
changes made previously will take effect.
For this example, configure channel 103 to measure a J-type
thermocouple with 0.1 °C of display resolution.
34970A Refresh UG.book Page 30 Wednesday, February 17, 2010 12:34 PM
Summary of Contents for 34970A
Page 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Page 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Page 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Page 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Page 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Page 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Page 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Page 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Page 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Page 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Page 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...