222
Chapter 5 Error Messages
Execution Errors
-214
Trigger deadlock
A trigger deadlock occurs when the trigger source is
“BUS”
and a
READ?
command is received.
-221
Settings conflict
An invalid configuration was requested. This error is most commonly
generated when setting alarm limits. Note that the lower limit must
always be less than or equal to the upper limit, even if you are using only
one of the limits. This error is also generated if you send the
MEASure?
or
CONFigure
command with autorange enabled with a fixed resolution.
-222
Data out of range
A numeric parameter value is outside the valid range for this command.
Example
:
TRIG:COUNT -3
-223
Too much data
A character string was received but could not be executed because the
string length was more than 12 characters. This error can be generated
by the
CAL:STRing
and
DISPlay:TEXT
commands.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for this
command. You may have used an invalid parameter choice.
Example
:
TRIG:SOURCE ALARM
(
ALARM
is not a valid choice)
-230
Data stale
A
FETCh?
or
DATA:REMove?
command was received but internal
reading memory was empty. The readings retrieved may be invalid.
-310
System error
A firmware defect has been found.
This is not a fatal error but you should
contact your nearest Agilent Service Center if this error is reported.
34970A Refresh UG.book Page 222 Wednesday, February 17, 2010 12:34 PM
Summary of Contents for 34970A
Page 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Page 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Page 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Page 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Page 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Page 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Page 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Page 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Page 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Page 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Page 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...