Chapter 7 Tutorial
Measurement Fundamentals
4
7
Loading Errors Due to Input Resistance
Measurement loading
errors occur when the resistance of the device-under-test (DUT) is an
appreciable percentage of the instrument’s own input resistance. The
diagram below shows this error source.
Where:
V
s
= Ideal DUT voltage
R
s
= DUT source resistance
R
i
= Input resistance (10 M
or > 10 G
)
Error (%) =
To minimize loading errors, set the DMM’s DC input resistance to
greater than 10 G
when needed (
for more information on DC input
resistance, see page 130
).
R
s
DMM
R
i
HI
LO
V
s
100
–
Rs
Rs Ri
+
---------------------------
34970A Refresh UG.book Page 279 Wednesday, February 17, 2010 12:34 PM
Summary of Contents for 34970A
Page 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Page 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Page 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Page 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Page 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Page 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Page 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Page 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Page 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Page 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Page 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...