114
Chapter 4 Features and Functions
Scanning with External Instruments
•
Remote Interface Operation
: The following program segment
configures the instrument for an externally controlled scan.
TRIG:SOUR TIMER
Select the scan interval
ROUT:CHAN:ADV:SOUR EXT
Select the channel advance source
TRIG:TIMER 5
Set the scan interval to 5 seconds
TRIG:COUNT 2
S
weep the scan list 2 times
INIT
Initiate the scan
To configure the instrument for 4-wire external scanning, send the
following command.
ROUTe:CHANnel:FWIRe {OFF|ON}[,(@<
ch_list
>)]
34970A Refresh UG.book Page 114 Wednesday, February 17, 2010 12:34 PM
Summary of Contents for 34970A
Page 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Page 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Page 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Page 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Page 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Page 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Page 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Page 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Page 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Page 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Page 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...