285
Chapter 7 Tutorial
Measurement Fundamentals
4
7
Calculate the approximate measurement error for a pulse train input
with a crest factor of 3 and a fundamental frequency of 20 kHz. The
internal DMM is set to the 1 V range. For this example, use the
90-day
accuracy specifications of ± (0.05% of r 0.04% of range), as shown
in chapter 8.
Error
sine
= ±(0.05% + 0.04%) = ±0.09%
Error
crest factor
= 0.15%
Error
bandwidth
=
* 100 = 1.4%
Total Error = 0.09% + 0.15% + 1.4% = 1.6%
Example: Calculating Measurement Error
3
–
2
20000
4 3.14159 1000000
-------------------------------------------------------------
34970A Refresh UG.book Page 285 Wednesday, February 17, 2010 12:34 PM
Summary of Contents for 34970A
Page 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Page 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Page 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Page 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Page 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Page 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Page 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Page 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Page 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Page 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Page 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...