195
Chapter 4 Features and Functions
Calibration Overview
4
Calibration Message
The instrument allows you to store one message in calibration memory
in the mainframe. For example, you can store such information as the
date when the last calibration was performed, the date when the next
calibration is due, the instrument’s serial number, or even the name and
phone number of the person to contact for a new calibration.
• You can record a calibration message
only
from the remote interface
and
only
when the instrument is unsecured. You can read the
message from either the front-panel or over the remote interface. You
can read the calibration message whether the instrument is secured
or unsecured.
• The calibration message may contain up to 40 characters. From the
front panel, you can view 13 characters of the message at a time.
Press
to scroll through the text of the message. Press
again to
increase the scrolling speed.
• Storing a calibration message will overwrite any message previously
stored in memory.
• The calibration message is stored in
non-volatile
memory in the
mainframe, and
does not
change when power has been off, after a
Factory Reset (
*RST
command), or after an Instrument Preset
(
SYSTem:PRESet
command).
•
Front-Panel Operation
:
CAL MESSAGE
•
Remote Interface Operation
: To store the calibration message, send
the following command.
CAL:STRING ’CAL: 06-01-98’
34970A Refresh UG.book Page 195 Wednesday, February 17, 2010 12:34 PM
Summary of Contents for 34970A
Page 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Page 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Page 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Page 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Page 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Page 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Page 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Page 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Page 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Page 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Page 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...