136
Chapter 4 Features and Functions
Mx+B Scaling
Mx+B Scaling
The scaling function allows you to apply a
gain
and
offset
to all readings
on a specified multiplexer channel during a scan. In addition to setting
the gain (“M”) and offset (“B”) values, you can also specify a custom
measurement label for your scaled readings (RPM, PSI, etc.). You can
apply scaling to any multiplexer channels and for any measurement
function. Scaling
is not
allowed with any of the digital channels on the
multifunction module.
• Scaling is applied using the following equation:
Scaled Reading = (Gain x Measurement) + Offset
• You must configure the channel (function, transducer type, etc.)
before applying any scaling values. If you change the measurement
configuration, scaling is turned off on that channel and the gain and
offset values are reset (M=1 and B=0). Scaling is also turned off when
you change the temperature probe type, temperature units, or disable
the internal DMM.
• If you plan to use scaling on a channel which will also use alarms,
be
sure to configure the scaling values first
. If you attempt to assign the
alarm limits first, the instrument will turn off alarms and clear the
limit values when you enable scaling on that channel. If you specify a
custom measurement label with scaling, it is automatically used
when alarms are logged on that channel.
• If you remove a channel from the scan list (by selecting
CHANNEL OFF
from the front panel or by redefining the scan list
from the remote interface), scaling is turned off for that channel but
the gain and offset values
are not
cleared. If you decide to add that
channel back to the scan list (without changing the function), the
original gain and offset values are restored and scaling is turned back
on. This makes it easy to temporarily remove a channel from the scan
list without entering the scaling values again.
34970A Refresh UG.book Page 136 Wednesday, February 17, 2010 12:34 PM
Summary of Contents for 34970A
Page 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Page 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Page 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Page 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Page 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Page 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Page 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Page 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Page 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Page 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Page 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...