131
Chapter 4 Features and Functions
Voltage Measurement Configuration
4
•
Remote Interface Operation
: You can enable or disable the automatic
input resistance mode on the specified channels. With AUTO OFF
(default), the input resistance is fixed at 10 M
for all ranges. With
AUTO ON, the input resistance is set to >10 G
for the three lowest
DC voltage ranges. The
MEASure?
and
CONFigure
commands
automatically select
AUTO OFF
.
INPUT:IMPEDANCE:AUTO ON,(@103)
AC Low Frequency Filter
The instrument uses three different AC filters which enable you to either
optimize low frequency accuracy or achieve faster AC settling times. The
instrument selects the
slow
,
medium
, or
fast
filter based on the input
frequency that you specify for the selected channels.
Applies to AC voltage and AC current measurements only.
• The instrument selects the
medium
filter (20 Hz) when the function is
changed or after a Factory Reset (
*RST
command). An Instrument
Preset (
SYSTem:PRESet
command) or Card Reset (
SYSTem:CPON
command)
does not
change the setting.
•
Front-Panel Operation
: First, select the AC voltage (or AC current)
function on the active channel. Then, go to the
Advanced
menu and
select the slow filter (3 Hz), medium filter (
20 Hz
), or fast filter (200
Hz) for the active channel.
The default is the medium filter.
LF 3 HZ:SLOW
•
Remote Interface Operation
: Specify the lowest frequency expected in
the input signal on the specified channels. The instrument selects the
appropriate filter based on the frequency you specify (
see table above
).
The
MEASure?
and
CONFigure
commands automatically select the
20 Hz (medium) filter.
SENS:VOLT:AC:BAND 3,(@203)
Input Frequency
Default Settling Delay
Minimum Settling Delay
3 Hz to 300 kHz (Slow)
20Hz to 300 kHz (Medium)
200 Hz to 300 kHz (Fast)
7 seconds / reading
1 second / reading
0.12 seconds / reading
1.5 seconds
0.2 seconds
0.02 seconds
Select the slow filter (3 Hz)
34970A Refresh UG.book Page 131 Wednesday, February 17, 2010 12:34 PM
Summary of Contents for 34970A
Page 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Page 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Page 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Page 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Page 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Page 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Page 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Page 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Page 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Page 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Page 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...